Dissolution tester
Contact Info
- Building 1, No. 5500 Yuanjiang Road, Minhang District, Shanghai,
- 高洁
- 13761436048
- 3982822529@qq.com
Dissolution Tester is a specialized pharmaceutical testing instrument used to detect the dissolution rate of solid preparations (such as tablets, capsules, etc.). It can simulate the gastrointestinal digestion process of the human body, and when combined with an ultraviolet spectrophotometer, it can detect the dissolution rate of pharmaceutical preparations.
This is an in vitro testing method for controlling the quality of pharmaceutical preparations, and it has been widely used in drug research, production, and testing. The improved dissolution tester adopts an imported motor, ensuring accurate rotation speed and stable operation. The instrument consists of components including a cabinet, control system, water bath system, transmission system, and built-in sensors. The precision control system composed of a single-chip microcomputer centrally controls all components; the instrument features high automation, precise control, small error, low noise, and convenient operation. The cabinet of the dissolution tester is composed of a machine head, electrical box, front control panel, water bath box, and other parts.
Main technical parameters of the dissolution tester are as follows:
I. Paddle shafts and basket shafts (collectively referred to as rotating shafts)
l 1. Specifications and dimensions comply with the regulations of the Chinese Pharmacopoeia
l 2. Paddle shaft swing amplitude: ≤±0.5mm
l 3. Rotating basket swing amplitude: ≤±1mm
l 4. Coaxiality between rotating shaft and dissolution cup: ≤2mm
II. Rotating shaft speed control
l 1. Speed adjustment range: 25~200 rpm
l 2. Speed resolution: 1 rpm
l 3. Speed stability error: ≤±4%
III. Water bath temperature control
l 1. Temperature adjustment range: 5.0 (room temperature) ~45.0℃
l 2. Temperature resolution: 0.1℃
l 3. Temperature control error: ≤±0.3℃
IV. Test time control
l 1. Total accumulated timing: Up to 99 hours and 59 minutes, counting up
l 2. Number of sampling cycles: Up to 9 different sampling cycles
l 3. Sampling cycle time: Up to 9 hours and 59 minutes per cycle, counting down
l 4. Timed on/off time: Up to 99 hours and 59 minutes, counting down
V. Operating noise: <60dB
VI. Continuous working time: 48 hours
VII. Safety performance: Complies with the national professional standard JB/T20076-2005, with self-test and automatic protection functions
VIII. Power supply: AC200V±10%, 50Hz, 1300W
IX. Environmental conditions: Ambient temperature 5~35℃, relative humidity ≤80%
| Instruments & Meters/Specialized Instruments/Asphalt Testing Instruments | |